ABSTRACT
The aim of this proposal is to design, develop, and validate an interactive easy-to-use computer modeling
tool that will help choose the best Transcatheter Pulmonary Valve Replacement (TPVR) device for an
individual patient. Increased availability of a variety of self-expanding TPVR devices with varying shapes and
sizes increases the number of candidates for TPVR but will require tools that can improve the efficiency and
accuracy of determining patient eligibility as well as the optimal, patient-specific device. Current screening
techniques are labor intensive, expensive, and continue to rely primarily on basic 2D morphological measures
from CT reconstructions of the right ventricular outflow tract (RVOT) and implantation of actual devices into 3D
printed models to visually assess device fit. Therefore, a clinical-grade application warrants integration of
realistic physics of device mechanics, deformable vessel walls and their interaction at higher accuracy, as well
as generation of quantitative metrics and visualizations within an intuitive user interface.
The proposed tool will model the vessel wall, self-centering of the self-expanding device and their
interactions using finite elements which will be incorporated as an external module for 3D Slicer-an open-
source medical imaging platform. Metrics such as device dimensions, compression, maximum stress from
contact, and assessment of proximal and distal circumferential seal will be automatically computed, visualized
and graphed for the cardiologists, facilitating determination of patient candidacy and the optimal choice of a
device for an individual patient. Our tool improves upon the current methods by incorporating (i) high-fidelity
contact modeling including a frictional contact model between the device and the vessel wall, (ii) automatic
estimation of the crucial metrics, (iii) near real-time performance allowing for incorporation into clinical
workflows, and (iv) methodology verification and clinical validation.
Modeling accuracy will be verified by comparing models based on existing pre-implant CT scans to post
device implant CT scans in an ovine model of pulmonary insufficiency. Further, the 3D Slicer extension tool will
be evaluated by three experienced interventional cardiologists (> 5 years of experience) at CHOP, who
routinely perform TPVR implant for its (1) ease of navigation; (2) quantitative metrics; (3) visualizations, and (4)
overall utility for clinical decision making.